Session Information
Date: Tuesday, November 10, 2026
Title: (2129-2151) Vasculitis - ANCA-Associated - Clinical Poster C
Session Type: Poster Session C
Session Time: 10:30AM-12:30PM
Background/Purpose: Autoantibody testing is central to diagnosing glomerular basement membrane (GBM) disease and ANCA-associated vasculitides (AAV). Key markers include IgG autoantibodies to myeloperoxidase (MPO), proteinase 3 (PR3) and GBM. Sequential or stepwise single-marker testing approaches may delay diagnosis, particularly in dual-positive cases, and increase laboratory workload. We evaluated a fully automated multiplex microarray immunoassay (MosaiQ AiPlex® VAS; AliveDx; Switzerland; investigational device) for simultaneous detection of anti-PR3, anti-MPO and anti-GBM IgG in two clinical laboratories in the Netherlands.
Methods: De-identified remnant human sera, pre-characterized for anti-PR3, anti-MPO and anti-GBM, were tested with the investigational device. Performance was compared with routine single-analyte assays [chemiluminescent immunoassay (CLIA) and fluorescent enzyme immunoassay (FEIA)]. Agreement was assessed using positive, negative and overall percent agreement (PPA, NPA, OPA) and Cohen’s kappa.
Results: 271 reactive (anti-MPO: 118; anti-PR3: 116; anti-GBM: 37) and 601 non-reactive results (anti-MPO: 204; anti-PR3: 202; anti-GBM: 197) were included. The investigational device showed strong agreement with the comparators (OPA across markers and methods: 96%, Cohen’s kappa: 0.91). Pooled PPA and NPA were 98.9% (95% CI: 96.8-99.8) and 94.7% (95% CI: 92.6-96.2), respectively. Additional details are shown in Table 1. Equivocal and invalid results rates were 0.1% and 0.2%.
Discordant samples were reviewed using available clinical information.
- Anti‑MPO: six samples were positive with the investigational device, but negative by both FEIA and CLIA, and two samples were positive with the investigational device and FEIA but negative by CLIA. All six with clinical data had a history of AAV, suggesting detection of persistent anti-MPO antibodies.
- Anti-PR3: eight samples were positive with the investigational device but negative by FEIA, four of which were also negative by CLIA. Four additional samples were negative by CLIA and positive with the investigational device and FEIA. Clinical data (n=4) indicated prior AAV. Three samples were negative with the investigational device but positive with the comparators, with inconsistent results across assays.
- Anti-GBM: two samples were positive with the investigational device but negative by FEIA; CLIA was positive in one case, and the other was not tested with the alternate comparator.
Further interpretation of discordant results was limited by incomplete clinical and analytical data.
Conclusion: The investigational multiplex microarray immunoassay demonstrated high agreement with routine single-marker assays and has the potential to provide an efficient approach for simultaneous detection of IgG autoantibodies to PR3, MPO and GBM. Its automated multiplex format may simplify serological testing pathways, while reducing testing burden. Further investigation of these potential benefits is warranted. These findings, consistent with previous results of the investigational device, support its potential utility in the laboratory evaluation of AAV and anti-GBM disease.
Table 1. Agreement of the investigational microarray with comparators
To cite this abstract in AMA style:
van den Bijgaart R, Westerhof P, Wetzels L, Namet I, Gomez G, Kamphuis S, van Suijlen J, Wevers B. Analytical Evaluation of a Fully Automated Multiplexed Microarray Immunoassay for Detection of Anti-PR3, Anti-MPO, and Anti-GBM Autoantibodies Compared with Routine Immunoassays in Two Clinical Laboratories in the Netherlands [abstract]. Arthritis Rheumatol. 2026; 78 (suppl 9). https://acrabstracts.org/abstract/analytical-evaluation-of-a-fully-automated-multiplexed-microarray-immunoassay-for-detection-of-anti-pr3-anti-mpo-and-anti-gbm-autoantibodies-compared-with-routine-immunoassays-in-two-clinical-labora/. Accessed .ACR Meeting Abstracts - https://acrabstracts.org/abstract/analytical-evaluation-of-a-fully-automated-multiplexed-microarray-immunoassay-for-detection-of-anti-pr3-anti-mpo-and-anti-gbm-autoantibodies-compared-with-routine-immunoassays-in-two-clinical-labora/