ACR Meeting Abstracts

ACR Meeting Abstracts

  • Meeting Abstracts
    • All Meetings
    • ACR Convergence 2026
    • PRSYM 2026
    • ACR Convergence 2025
    • Download Abstract Supplements
  • Keyword Index
  • My Favorites
    • View & print my favorites
  • Search

Home » Meeting Abstracts » ACR Convergence 2026

Abstract Number: 2130

Analytical Evaluation of a Fully Automated Multiplexed Microarray Immunoassay for Detection of Anti-PR3, Anti-MPO, and Anti-GBM Autoantibodies Compared with Routine Immunoassays in Two Clinical Laboratories in the Netherlands

Renske J.E. van den Bijgaart1, Puck A. Westerhof1, Lisanne S. Wetzels2, Inana Namet3, Gerber Gomez4, Stephan Kamphuis1, Jeroen D.E. van Suijlen1 and Brigitte A. Wevers2

Affiliations: 1Laboratory of Clinical Chemistry and Hematology, Clinical Diagnostics Gelre, Apeldoorn, Netherlands, 2Division of Laboratories, Pharmacy and Biomedical Genetics, Central Diagnostic Laboratory, University Medical Center Utrecht, Utrecht, Netherlands, 3Product Management, AliveDx Suisse SA, Eysins, Switzerland, 4Scientific and Medical Affairs, AliveDx Suisse SA, Eysins, Vaud, Switzerland

Share on X (Twitter) Share on Facebook Share on LinkedIn Share on Email
Print

Meeting: ACR Convergence 2026

Keywords: ANCA associated vasculitis, Anti-GBM Disease (Goodpasture's), Autoantibody(ies), autoantigens, Biomarkers

Session Information

Date: Tuesday, November 10, 2026

Title: (2129-2151) Vasculitis - ANCA-Associated - Clinical Poster C

Session Type: Poster Session C

Session Time: 10:30AM-12:30PM

Background/Purpose: Autoantibody testing is central to diagnosing glomerular basement membrane (GBM) disease and ANCA-associated vasculitides (AAV). Key markers include IgG autoantibodies to myeloperoxidase (MPO), proteinase 3 (PR3) and GBM. Sequential or stepwise single-marker testing approaches may delay diagnosis, particularly in dual-positive cases, and increase laboratory workload. We evaluated a fully automated multiplex microarray immunoassay (MosaiQ AiPlex® VAS; AliveDx; Switzerland; investigational device) for simultaneous detection of anti-PR3, anti-MPO and anti-GBM IgG in two clinical laboratories in the Netherlands.

Methods: De-identified remnant human sera, pre-characterized for anti-PR3, anti-MPO and anti-GBM, were tested with the investigational device. Performance was compared with routine single-analyte assays [chemiluminescent immunoassay (CLIA) and fluorescent enzyme immunoassay (FEIA)]. Agreement was assessed using positive, negative and overall percent agreement (PPA, NPA, OPA) and Cohen’s kappa.

Results: 271 reactive (anti-MPO: 118; anti-PR3: 116; anti-GBM: 37) and 601 non-reactive results (anti-MPO: 204; anti-PR3: 202; anti-GBM: 197) were included. The investigational device showed strong agreement with the comparators (OPA across markers and methods: 96%, Cohen’s kappa: 0.91). Pooled PPA and NPA were 98.9% (95% CI: 96.8-99.8) and 94.7% (95% CI: 92.6-96.2), respectively. Additional details are shown in Table 1. Equivocal and invalid results rates were 0.1% and 0.2%.

Discordant samples were reviewed using available clinical information.

  • Anti‑MPO: six samples were positive with the investigational device, but negative by both FEIA and CLIA, and two samples were positive with the investigational device and FEIA but negative by CLIA. All six with clinical data had a history of AAV, suggesting detection of persistent anti-MPO antibodies.
  • Anti-PR3: eight samples were positive with the investigational device but negative by FEIA, four of which were also negative by CLIA. Four additional samples were negative by CLIA and positive with the investigational device and FEIA. Clinical data (n=4) indicated prior AAV. Three samples were negative with the investigational device but positive with the comparators, with inconsistent results across assays.
  • Anti-GBM: two samples were positive with the investigational device but negative by FEIA; CLIA was positive in one case, and the other was not tested with the alternate comparator.

Further interpretation of discordant results was limited by incomplete clinical and analytical data.

Conclusion: The investigational multiplex microarray immunoassay demonstrated high agreement with routine single-marker assays and has the potential to provide an efficient approach for simultaneous detection of IgG autoantibodies to PR3, MPO and GBM. Its automated multiplex format may simplify serological testing pathways, while reducing testing burden. Further investigation of these potential benefits is warranted. These findings, consistent with previous results of the investigational device, support its potential utility in the laboratory evaluation of AAV and anti-GBM disease.

Supporting image 1Table 1. Agreement of the investigational microarray with comparators


Disclosures: R. van den Bijgaart: None; P. Westerhof: None; L. Wetzels: None; I. Namet: AliveDx Suisse SA, 12,; G. Gomez: AliveDx, 12,; S. Kamphuis: None; J. van Suijlen: None; B. Wevers: None.

To cite this abstract in AMA style:

van den Bijgaart R, Westerhof P, Wetzels L, Namet I, Gomez G, Kamphuis S, van Suijlen J, Wevers B. Analytical Evaluation of a Fully Automated Multiplexed Microarray Immunoassay for Detection of Anti-PR3, Anti-MPO, and Anti-GBM Autoantibodies Compared with Routine Immunoassays in Two Clinical Laboratories in the Netherlands [abstract]. Arthritis Rheumatol. 2026; 78 (suppl 9). https://acrabstracts.org/abstract/analytical-evaluation-of-a-fully-automated-multiplexed-microarray-immunoassay-for-detection-of-anti-pr3-anti-mpo-and-anti-gbm-autoantibodies-compared-with-routine-immunoassays-in-two-clinical-labora/. Accessed .

ACR Meeting Abstracts - https://acrabstracts.org/abstract/analytical-evaluation-of-a-fully-automated-multiplexed-microarray-immunoassay-for-detection-of-anti-pr3-anti-mpo-and-anti-gbm-autoantibodies-compared-with-routine-immunoassays-in-two-clinical-labora/

My Favorites

Save and print abstracts during your browser session by clicking the “Favorite” button at the bottom of any abstract (must have cookies enabled in browser). See saved favorites.

Abstract Policies

  • ACR Convergence Abstract Embargo Policies
  • ACR Convergence Abstract Permissions & Reprints
  • PRSYM Abstract Policies

ACR Convergence. Where Rheumatology Meets

ACR Convergence 2026

Join us November 6-11 in Orlando, Florida.
See Registration Information

  • Contact ACR
  • Privacy Policy
  • ACR Policies
  • Cookie Preferences

© Copyright 2026 American College of Rheumatology